International Journal of Radiation Oncology * Biology * Physics
Volume 75, Issue 1 , Pages 40-48, 1 September 2009

Dosimetric Changes Resulting From Patient Rotational Setup Errors in Proton Therapy Prostate Plans

  • Samir V. Sejpal, M.D., M.P.H.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Richard A. Amos, M.S.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Jaques B. Bluett, M.S.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Lawrence B. Levy, M.S.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Rajat J. Kudchadker, Ph.D.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Jennifer Johnson, M.S.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Seungtaek Choi, M.D.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
  • ,
  • Andrew K. Lee, M.D., M.P.H.

      Affiliations

    • Division of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, Houston, TX
    • Corresponding Author InformationReprint requests to: Andrew K. Lee, M.D., M.P.H., Department of Radiation Oncology, University of Texas M. D. Anderson Cancer Center, 1515 Holcombe Blvd., Unit 1202, Houston, TX 77030-4009. Tel: (713) 563-2348; Fax: (713) 563-6940

Received 3 March 2008; received in revised form 26 August 2008; accepted 26 August 2008. published online 05 December 2008.

Purpose

To evaluate the dose changes to the target and critical structures from rotational setup errors in prostate cancer patients treated with proton therapy.

Methods and Materials

A total of 70 plans were analyzed for 10 patients treated with parallel-opposed proton beams to a dose of 7,600 60Co-cGy-equivalent (CcGE) in 200 CcGE fractions to the clinical target volume (i.e., prostate and proximal seminal vesicles). Rotational setup errors of +3°, −3°, +5°, and −5° (to simulate pelvic tilt) were generated by adjusting the gantry. Horizontal couch shifts of +3° and −3° (to simulate longitudinal setup variability) were also generated. Verification plans were recomputed, keeping the same treatment parameters as the control.

Results

All changes shown are for 38 fractions. The mean clinical target volume dose was 7,780 CcGE. The mean change in the clinical target volume dose in the worse case scenario for all shifts was 2 CcGE (absolute range in worst case scenario, 7,729–7,848 CcGE). The mean changes in the critical organ dose in the worst case scenario was 6 CcGE (bladder), 18 CcGE (rectum), 36 CcGE (anterior rectal wall), and 141 CcGE (femoral heads) for all plans. In general, the percentage of change in the worse case scenario for all shifts to the critical structures was <5%. Deviations in the absolute percentage of volume of organ receiving 45 and 70 Gy for the bladder and rectum were <2% for all plans.

Conclusion

Patient rotational movements of 3° and 5° and horizontal couch shifts of 3° in prostate proton planning did not confer clinically significant dose changes to the target volumes or critical structures.

Proton therapy, Prostate cancer, Rotational setup errors, Dosimetry, Treatment planning

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 30.00 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Conflict of interest: none.

PII: S0360-3016(08)03431-7

doi:10.1016/j.ijrobp.2008.08.042

International Journal of Radiation Oncology * Biology * Physics
Volume 75, Issue 1 , Pages 40-48, 1 September 2009