Quantifying Interfraction and Intrafraction Tumor Motion in Lung Stereotactic Body Radiotherapy Using Respiration-Correlated Cone Beam Computed Tomography

  • Jean-Pierre Bissonnette, Ph.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
    • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada
    • Corresponding Author InformationReprint requests to: Jean-Pierre Bissonnette, Ph.D., Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada. Tel: (416) 946-4501, ext. 2151; Fax: (416) 946-5655
  • ,
  • Kevin N. Franks, M.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
  • ,
  • Thomas G. Purdie, Ph.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
    • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada
  • ,
  • Douglas J. Moseley, Ph.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
    • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada
  • ,
  • Jan-Jakob Sonke, Ph.D.

      Affiliations

    • Department of Radiation Oncology, The Netherlands Cancer Institute, Antoni van Leeuwenhoek Hospital, Amsterdam, The Netherlands
  • ,
  • David A. Jaffray, Ph.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
    • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada
  • ,
  • Laura A. Dawson, M.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
    • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada
  • ,
  • Andrea Bezjak, M.D.

      Affiliations

    • Radiation Medicine Program, Princess Margaret Hospital, University Health Network, Toronto, Ontario, Canada
    • Department of Radiation Oncology, University of Toronto, Toronto, Ontario, Canada

Received 25 July 2008 ,Revised 7 November 2008 ,Accepted 18 November 2008.

References 

  1. International Commission on Radiation Units and Measurements. ICRU report 62: Prescribing, recording, and reporting photon beam therapy (supplement to ICRU Report 50). Bethesda, MD: ICRU; 1999;
  2. Siewerdsen JH, Jaffray DA. A ghost story: spatio-temporal response characteristics of an indirect-detection flat-panel imager. Med Phys. 1999;26:1624–1641
  3. Blomgren H, Lax I, Naslund I, et al. Stereotactic high dose fraction radiation therapy of extracranial tumors using an accelerator. Clinical experience of the first thirty-one patients. Acta Oncol. 1995;34:861–870
  4. Wulf J, Hadinger U, Oppitz U, et al. Stereotactic radiotherapy of extracranial targets: CT-simulation and accuracy of treatment in the stereotactic body frame. Radiother Oncol. 2000;57:225–236
  5. Wong JW, Sharpe MB, Jaffray DA, et al. The use of active breathing control (ABC) to reduce margin for breathing motion. Int J Radiat Oncol Biol Phys. 1999;44:911–919
  6. Kimura T, Hirokawa Y, Murakami Y, et al. Reproducibility of organ position using voluntary breath-hold method with spirometer for extracranial stereotactic radiotherapy. Int J Radiat Oncol Biol Phys. 2004;60:1307–1313
  7. Heinzerling JH, Anderson JF, Papiez L, et al. Four-dimensional computed tomography scan analysis of tumor and organ motion at varying levels of abdominal compression during stereotactic treatment of lung and liver. Int J Radiat Oncol Biol Phys. 2008;70:1571–1578
  8. Purdie TG, Bissonnette JP, Franks K, et al. Cone-beam computed tomography for on-line image guidance of lung stereotactic radiotherapy: localization, verification, and intrafraction tumor position. Int J Radiat Oncol Biol Phys. 2007;68:243–252
  9. Hof H, Herfarth KK, Munter M, et al. The use of the multislice CT for the determination of respiratory lung tumor movement in stereotactic single-dose irradiation. Strahlenther Onkol. 2003;179:542–547
  10. Whyte RI, Crownover R, Murphy MJ, et al. Stereotactic radiosurgery for lung tumors: preliminary report of a phase I trial. Ann Thorac Surg. 2003;75:1097–1101
  11. Shimizu S, Shirato H, Kitamura K, et al. Use of an implanted marker and real-time tracking of the marker for the positioning of prostate and bladder cancers. Int J Radiat Oncol Biol Phys. 2000;48:1591–1597
  12. Shirato H, Shimizu S, Kitamura K, et al. Four-dimensional treatment planning and fluoroscopic real-time tumor tracking radiotherapy for moving tumor. Int J Radiat Oncol Biol Phys. 2000;48:435–442
  13. Rietzel E, Pan T, Chen GT. Four-dimensional computed tomography: Image formation and clinical protocol. Med Phys. 2005;32:874–889
  14. Hugo G, Vargas C, Liang J, et al. Changes in the respiratory pattern during radiotherapy for cancer in the lung. Radiother Oncol. 2006;78:326–331
  15. Bosmans G, van Baardwijk A, Dekker A, et al. Intra-patient variability of tumor volume and tumor motion during conventionally fractionated radiotherapy for locally advanced non-small-cell lung cancer: A prospective clinical study. Int J Radiat Oncol Biol Phys. 2006;66:748–753
  16. Onimaru R, Fujino M, Yamazaki K, et al. Steep dose-response relationship for stage I non-small-cell lung cancer using hypofractionated high-dose irradiation by real-time tumor-tracking radiotherapy. Int J Radiat Oncol Biol Phys. 2008;70:374–381
  17. Sonke JJ, Zijp L, Remeijer P, et al. Respiratory correlated cone beam CT. Med Phys. 2005;32:1176–1186
  18. Purdie TG, Moseley DJ, Bissonnette JP, et al. Respiration correlated cone-beam computed tomography and 4DCT for evaluating target motion in stereotactic lung radiation therapy. Acta Oncol. 2006;45:915–922
  19. Mageras GS, Pevsner A, Yorke ED, et al. Measurement of lung tumor motion using respiration-correlated CT. Int J Radiat Oncol Biol Phys. 2004;60:933–941
  20. Lax I, Blomgren H, Naslund I, et al. Stereotactic radiotherapy of malignancies in the abdomen. Methodological aspects. Acta Oncol. 1994;33:677–683
  21. Blomgren H, Lax I, Goranson H, et al. Radiosurgery for tumours in the body: clinical experience using a new method. J Radiosurg. 1998;1:63–74
  22. McGarry RC, Papiez L, Williams M, et al. Stereotactic body radiation therapy of early-stage non-small-cell lung carcinoma: phase I study. Int J Radiat Oncol Biol Phys. 2005;63:1010–1015
  23. Feldkamp L, Davis L, Kress J. Practical cone-beam algorithm. Opt Soc Am A. 1984;1:612–619
  24. Shirato H, Shimizu S, Kunieda T, et al. Physical aspects of a real-time tumor-tracking system for gated radiotherapy. Int J Radiat Oncol Biol Phys. 2000;48:1187–1195
  25. Seppenwoolde Y, Shirato H, Kitamura K, et al. Precise and real-time measurement of 3D tumor motion in lung due to breathing and heartbeat, measured during radiotherapy. Int J Radiat Oncol Biol Phys. 2002;53:822–834
  26. Haasbeek CJ, Lagerwaard FJ, Cuijpers JP, et al. Is adaptive treatment planning required for stereotactic radiotherapy of stage I non-small-cell lung cancer?. Int J Radiat Oncol Biol Phys. 2007;67:1370–1374
  27. van der Geld YG, Lagerwaard FJ, van Sornsen de Koste JR, et al. Reproducibility of target volumes generated using uncoached 4-dimensional CT scans for peripheral lung cancer. Radiat Oncol. 2006;1:43
  28. Guckenberger M, Wilbert J, Meyer J, et al. Is a single respiratory correlated 4D-CT study sufficient for evaluation of breathing motion?. Int J Radiat Oncol Biol Phys. 2007;67:1352–1359
  29. Guckenberger M, Meyer J, Wilbert J, et al. Intra-fractional uncertainties in cone-beam CT based image-guided radiotherapy (IGRT) of pulmonary tumors. Radiother Oncol. 2007;83:57–64
  30. Sonke JJ, Lebesque J, van Herk M. Variability of four-dimensional computed tomography patient models. Int J Radiat Oncol Biol Phys. 2008;70:590–598
  31. Li XA, Stepaniak C, Gore E. Technical and dosimetric aspects of respiratory gating using a pressure-sensor motion monitoring system. Med Phys. 2006;33:145–154

 This research was supported by the Addie MacNaughton Chair in Thoracic Radiation Oncology and by Elekta Oncology Systems. Kevin Franks, M.D., is the recipient of a grant from the Canadian Institute of Health Research (CIHR) Strategic Training Fellowship in the Excellence in Radiation Research for the 21st Century (EIRR21) program.

 Conflict of interest: This research was supported in part by grants from Elekta Oncology Systems. The authors' institutions are members of the Elekta Synergy Research Group, together with the William Beaumont Hospital, Royal Oak, MI; Thomas Jefferson University, Philadelphia, PA; Christie Hospital, Manchester, UK; and Elekta Oncology Systems Ltd., Crawley, West Sussex, UK.

PII: S0360-3016(09)00194-1

doi: 10.1016/j.ijrobp.2008.11.066

International Journal of Radiation Oncology * Biology * Physics
Volume 75, Issue 3 , Pages 688-695 , 1 November 2009