Patient Motion and Targeting Accuracy in Robotic Spinal Radiosurgery: 260 Single-Fraction Fiducial-Free Cases

  • Christoph Fürweger, Ph.D.

      Affiliations

    • European Cyberknife Center, Munich, Germany
    • Corresponding Author InformationReprint requests to: Christoph Fürweger, Ph.D., Max-Lebsche-Platz 31, 81377 Munich, Germany. Tel: (+49) 89-452336-0; Fax: (+49) 89-452336-16
  • ,
  • Christian Drexler

      Affiliations

    • European Cyberknife Center, Munich, Germany
  • ,
  • Markus Kufeld, M.D.

      Affiliations

    • European Cyberknife Center, Munich, Germany
  • ,
  • Alexander Muacevic, M.D.

      Affiliations

    • European Cyberknife Center, Munich, Germany
  • ,
  • Berndt Wowra, M.D.

      Affiliations

    • European Cyberknife Center, Munich, Germany
  • ,
  • Alexander Schlaefer, Ph.D.

      Affiliations

    • University of Luebeck, Institute for Robotics and Cognitive Systems, Luebeck, Germany

Received 3 July 2009; received in revised form 19 October 2009; accepted 23 November 2009. published online 14 April 2010.

Purpose

To evaluate clinical targeting precision and assess patient movement data during fiducial-free, single-fraction spinal radiosurgery with the Cyberknife (CK).

Methods and Materials

Image-guided spine tracking accuracy was tested using two phantoms. Movement patterns (three translations, roll, pitch and yaw) were obtained from log files of 260 patient treatments (47 cervical, 89 thoracic, 90 lumbar, and 34 pelvic/sacral). For two treatments (average and maximum motion scenario), we added offsets to all beams according to recorded patient movements and recalculated the delivered dose distribution to simulate the dosimetric impact of intrafraction motion.

Results

Phantom spine position was registered with an accuracy of <0.2 mm for translational and <0.3° for rotational directions. Residual patient motion yielded mean targeting errors per beam of 0.28 ± 0.13 mm (X), 0.25 ± 0.15 mm (Y), 0.19 ± 0.11 mm (Z) and 0.40 ± 0.20° (roll), 0.20 ± 0.08° (pitch), and 0.19 ± 0.08° (yaw). Spine region had little influence on overall targeting error, which was <1 mm for more than 95% of treatments (median, 0.48 mm). In the maximum motion case, target coverage decreased by 1.7% (from 92.1% to 90.4%) for the 20-Gy prescription isodose. Spinal cord volume receiving more than 8 Gy increased slightly, from 2.41 to 2.46 cm3.

Conclusions

Submillimeter targeting precision was obtained for fiducial-free spinal radiosurgery despite patient motion. Patient motion has little effect on the delivered dose distribution when image-guided correction of beam aiming is employed.

Spinal radiosurgery, Spine, Fiducial free registration, Intrafraction motion, Cyberknife

To access this article, please choose from the options below

Login to an existing account or Register a new account.

  • Purchase this article for 30.00 USD (You must login/register to purchase this article)

    Online access for 24 hours. The PDF version can be downloaded as your permanent record.

  • Subscribe to this title

    Get unlimited online access to this article and all other articles in this title 24/7 for one year.

  • Claim access now

    For current subscribers with Society Membership or Account Number.

  • Visit SciVerse ScienceDirect to see if you have access via your institution.
 

 Conflict of interest: none.

PII: S0360-3016(09)03571-8

doi:10.1016/j.ijrobp.2009.11.030

International Journal of Radiation Oncology * Biology * Physics
Volume 78, Issue 3 , Pages 937-945, 1 November 2010