Experimental Platform for Ultra-high Dose Rate FLASH Irradiation of Small Animals Using a Clinical Linear Accelerator

Published:September 20, 2016DOI:https://doi.org/10.1016/j.ijrobp.2016.09.018

      Purpose

      A key factor limiting the effectiveness of radiation therapy is normal tissue toxicity, and recent preclinical data have shown that ultra-high dose rate irradiation (>50 Gy/s, “FLASH”) potentially mitigates this effect. However, research in this field has been strongly limited by the availability of FLASH irradiators suitable for small animal experiments. We present a simple methodologic approach for FLASH electron small animal irradiation with a clinically available linear accelerator (LINAC).

      Methods and Materials

      We investigated the FLASH irradiation potential of a Varian Clinac 21EX in both clinical mode and after tuning of the LINAC. We performed detailed FLUKA Monte Carlo and experimental dosimetric characterization at multiple experimental locations within the LINAC head.

      Results

      Average dose rates of ≤74 Gy/s were achieved in clinical mode, and the dose rate after tuning exceeded 900 Gy/s. We obtained 220 Gy/s at 1-cm depth for a >4-cm field size with 90% homogeneity throughout a 2-cm-thick volume.

      Conclusions

      We present an approach for using a clinical LINAC for FLASH irradiation. We obtained dose rates exceeding 200 Gy/s after simple tuning of the LINAC, with excellent dosimetric properties for small animal experiments. This will allow for increased availability of FLASH irradiation to the general research community.
      To read this article in full you will need to make a payment
      ASTRO Member Login
      ASTRO Members, full access to the journal is a member benefit. Use your society credentials to access all journal content and features.
      Already a print subscriber? Claim online access
      Already an online subscriber? Sign in
      Institutional Access: Sign in to ScienceDirect

      References

        • Lohse I.
        • Lang S.
        • Hrbacek J.
        • et al.
        Effect of high dose per pulse flattening filter-free beams on cancer cell survival.
        Radiother Oncol. 2011; 101: 226-232
        • Ling C.C.
        • Gerweck L.E.
        • Zaider M.
        • et al.
        Dose-rate effects in external beam radiotherapy redux.
        Radiother Oncol. 2010; 95: 261-268
        • Ponette V.
        • Le Péchoux C.
        • Deniaud-Alexandre E.
        • et al.
        Hyperfast, early cell response to ionizing radiation.
        Int J Radiat Biol. 2000; 76: 1233-1243
        • Auer S.
        • Hable V.
        • Greubel C.
        • et al.
        Survival of tumor cells after proton irradiation with ultra-high dose rates.
        Radiat Oncol. 2011; 6: 139
        • Zlobinskaya O.
        • Siebenwirth C.
        • Greubel C.
        • et al.
        The effects of ultra-high dose rate proton irradiation on growth delay in the treatment of human tumor xenografts in nude mice.
        Radiat Res. 2014; 181: 177-183
        • Matsuura T.
        • Egashira Y.
        • Nishio T.
        • et al.
        Apparent absence of a proton beam dose rate effect and possible differences in RBE between Bragg peak and plateau.
        Med Phys. 2010; 37: 5376-5381
        • Rafat M.
        • Bazalova M.
        • Palma B.
        • et al.
        SU-C-BRE-06: Radiobiological advantage of very rapid irradiation.
        Med Phys. 2014; 41: 95
        • Rafat M.
        • Bazalova M.
        • Palma B.A.
        • et al.
        Impact of very rapid irradiation on clonogenic survival.
        Int J Radiat Oncol Biol Phys. 2014; 90: S790
        • Favaudon V.
        • Caplier L.
        • Monceau V.
        • et al.
        Ultrahigh dose-rate flash irradiation increases the differential response between normal and tumor tissue in mice.
        Sci Transl Med. 2014; 6: 245ra293
        • Stojadinovic S.
        • Low D.
        • Hope A.
        • et al.
        MicroRT—Small animal conformal irradiator.
        Med Phys. 2007; 34: 4706-4716
        • Tryggestad E.
        • Armour E.
        • Deng H.
        • et al.
        WE-D-330A-08: The small-animal radiation research platform (SARRP): Commissioning a 225 kvp “small-field” x-ray source for Monte Carlo-based treatment planning.
        Med Phys. 2006; 33: 2241
        • Jaffray D.
        • Moseley D.
        • Chow J.
        • et al.
        WE-D-330A-09: An image-guided irradiator for pre-clinical radiation therapy studies.
        Med Phys. 2006; 33: 2241
        • Kiehl E.L.
        • Stojadinovic S.
        • Malinowski K.T.
        • et al.
        Feasibility of small animal cranial irradiation with the microRT system.
        Med Phys. 2008; 35: 4735-4743
        • Chao T.
        • Chen A.
        • Tu S.
        • et al.
        The evaluation of 6 and 18 mev electron beams for small animal irradiation.
        Phys Med Biol. 2009; 54: 5847-5860
        • Böhlen T.
        • Cerutti F.
        • Chin M.
        • et al.
        The FLUKA code: Developments and challenges for high energy and medical applications.
        Nucl Data Sheets. 2014; 120: 211-214
        • Ferrari A.
        • Sala P.R.
        • Fasso A.
        • et al.
        FLUKA: A Multi-particle Transport Code (Program Version 2005).
        2005 (No. INFN-TC-05-11) (Accessed June, 2015)
        • Thebaut J.
        Measurement Driven, Electron Beam Modeling and Commissioning for a Monte Carlo Treatment Planning System With Improved Accuracy.
        McGill Library and Collections, Montreél, Québec Canada2009 (Available at:) (Accessed June 2015)
        • Chiu-Tsao S.T.
        • Ho Y.
        • Shankar R.
        • et al.
        Energy dependence of response of new high sensitivity radiochromic films for megavoltage and kilovoltage radiation energies.
        Med Phys. 2005; 32: 3350-3354
        • Arjomandy B.
        • Tailor R.
        • Anand A.
        • et al.
        Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron, and proton beam energies.
        Med Phys. 2010; 37: 1942-1947
        • Bazalova-Carter M.
        • Liu M.
        • Palma B.
        • et al.
        Comparison of film measurements and Monte Carlo simulations of dose delivered with very high-energy electron beams in a polystyrene phantom.
        Med Phys. 2015; 42: 1606-1613
        • Podgorsak E.B.
        Radiation Physics for Medical Physicists, 2nd ed.
        in: Greenbaum E. Aizawa M. Andersen O.S. Biological and Medical Physics Biomedical Engineering. Springer-Verlag, Berlin2010
        • Ding G.X.
        Energy spectra, angular spread, fluence profiles and dose distributions of 6 and 18 mv photon beams: Results of Monte Carlo simulations for a Varian 2100EX accelerator.
        Phys Med Biol. 2002; 47: 1025-1046
        • Siebers J.V.
        • Keall P.J.
        • Libby B.
        • et al.
        Comparison of EGS4 and MCNP4B Monte Carlo codes for generation of photon phase space distributions for a Varian 2100C.
        Phys Med Biol. 1999; 44: 3009-3026
        • Rustgi S.N.
        • Working K.R.
        Dosimetry of small field electron beams.
        Med Dosim. 1992; 17: 107-110
        • Rashid H.
        • Islam M.K.
        • Gaballa H.
        • et al.
        Small-field electron dosimetry for the Philips SL25 linear accelerator.
        Med Phys. 1990; 17: 710-714
        • Favaudon V.
        • Tourbez H.
        • Houee-Levin C.
        • et al.
        Co2.- radical induced cleavage of disulfide bonds in proteins: A gamma-ray and pulse radiolysis mechanistic investigation.
        Biochemistry. 1990; 29: 10978-10989
        • Petersson K.
        • Jaccard M.
        • Vozenin M.
        • et al.
        Dosimetry of ultra high dose rate irradiation for studies on the biological effect induced in normal brain and GBM.
        Radiother Oncol. 2016; 118: S84
        • Bazalova-Carter M.
        • Qu B.
        • Palma B.
        • et al.
        Treatment planning for radiotherapy with very high-energy electron beams and comparison of VHEE and VMAT plans.
        Med Phys. 2015; 42: 2615-2625
        • Palma B.
        • Bazalova-Carter M.
        • Hardemark B.
        • et al.
        Assessment of the quality of very high-energy electron radiotherapy planning.
        Radiother Oncol. 2016; 119: 154-158

      Comments

      Commenting Guidelines

      To submit a comment for a journal article, please use the space above and note the following:

      • We will review submitted comments as soon as possible, striving for within two business days.
      • This forum is intended for constructive dialogue. Comments that are commercial or promotional in nature, pertain to specific medical cases, are not relevant to the article for which they have been submitted, or are otherwise inappropriate will not be posted.
      • We require that commenters identify themselves with names and affiliations.
      • Comments must be in compliance with our Terms & Conditions.
      • Comments are not peer-reviewed.